| Pressure Related help: XN-L |
| Error Code |
-
119010: Abnormal Pressure loss
-
113030: 0.06MPa pressure error
|
| Symptom |
- Error message has come up on the IPU.
- Error message flickers on the IPU.
- Operation of the analyser are not working correctly.
- Sampling analysis errors.
|
| Possible Causes |
- The 0.06 MPa relief valve has malfunctioned.
- The air pump has malfunctioned.
|
| Refining Problem Area |
- Which pressure error is visible?
- How frequent is the error occurring?
- Can you hear any abnormal hissing noises or leaks?
- What function of the analyser have been affected?
|
| Action |
- Click [Execute] in the dialogue box. An air pump test will be executed. Record highest and lowest 0.06MPa pressure reached during the air pump test.
- Note if the air pump test has passed or failed.
|