| Cassette Related Errors – SP:10 |
| Error Code |
- 424100.0.0: No Cassette
- 424101.0.0: Supply cassette
- 247010.0.0: No Cassette shift home pos.
- 247160.0.0: No Cassette at storage pos.
- 424110.0.0: Remove cassette from table.
|
| Symptom |
- Operation of the analyser has stopped.
- Cassettes not moving into correct locations.
- Cassettes not moving in the stain table.
- Cassettes not moving in cassette storage location.
|
| Possible Causes |
- There is no cassette on the cassette supply table.
- The cassette is not placed correctly on the cassette supply table.
- There are 10 or less cassettes on the cassette supply table.
- There is a foreign object in the past of the cassette.
- There is a foreign object in the cassette shift unit.
- There are 70 or more cassettes on the cassette feed out table.
|
| Refining Problem Area |
- Which location of the analyser is the cassette not moving correctly?
- Where is the noise or vibration coming from?
- Is the error occurring frequently or intermittently?
- Is the error occurring after every cassette?
- Which stain pipette is the error occurring at?
|
| Action |
- Remove the cassettes and then follow the instructions in the action message to reset.
- Remove foreign object, then follow instruction in the action message to reset.
- Fill cassette supply table.
- Remove cassettes from cassette feed out table.
|