| Blood Aspiration Related Errors: XN-10/20 |
| Error Code |
- 161200, 161010: Insufficient blood volume.
- 161000: Blood cannot be aspirated.
- 161100: Aspiration Sensor Error.
|
| Symptom |
- Multiple aspiration errors.
- Inconsistent patient sample results.
- Error is not executable.
|
| Possible Causes |
- Density of the sample is inconsistent.
- Piercer or tubing of WB aspiration line is clogged.
- Blood aspiration sensor is malfunctioned.
- Blood sensor could not detect the specified amount of blood.
- Blood volume is insufficient.
|
| Refining Problem Area |
- Is the error occurring with every patient sample?
- Have you tried running the patient sample again?
- Does this error occur with QC and patient samples?
- Are there any fluid leaks around the analyser?
|
| Action |
- Re-mix the sample well then re-analyse.
- Perform [Auto rinse] procedure then analyse sample.
- Repeat the analysis by manual or micro analysis.
|