| Reagent Related Errors |
| Error Code |
- 000043: Out of Cellpack DCL
- 000044: Out of Stain1 Solution
- 000049: Out of phosphate buffer.
- 000046: Out of Rinse water
- 000050: Out of Methanol.
|
| Symptom |
- Error message is not executable.
- Stain 1 Chamber not filling.
- Phosphate buffer chamber not filling.
- Less than optimal slide quality.
- Analyser not able to perform startup sequence.
|
| Possible Causes |
- Reagent container is empty.
- There is a kink or bend in the reagent tubing.
- There is a blockage in the reagent tubing.
- Reagent tubing is not connected properly.
|
| Refining Problem Area |
- How often is the error occurring?
- Does the same error often or are there different errors?
- Are any other function of the analyser affected?
- Are there any noises or leaks present?
|
| Action |
- Inspect intake tubing from reagent bottle to analyser. Ensure there are no kinks or bends in the tubing.
- Inspect intake tubing from reagent bottle to analyser. Ensure there are no obvious blockages.
- Replace reagent and execute error.
Reagent Replacement Procedure
- Prepare the new reagent.
- Touch [Execute] in the [Help] dialog box.
- Input the reagent information of new regent.
- Select appropriate reagent then input the following information.
- Cellpack DCL: Replace bottle, Scan reagent code. - Methanol: Replace bottle. - Phosphate buffer: Replace bottle. - Stain 1: Replace bottle. - Rinse water: Replace bottle.

|